3D Printing as an Adaptive Tool for Sensor Fabrication Applied through pH Monitoring

01

Summary

What the paper reports, in its own words.

AI Enhanced· 72%

Abstract

As a key parameter, pH has received a lot of attention from the sensing perspective. New materials and technologies are being used to produce state of the art devices capable of tracking it. However, most generic sensors lack the applicability that certain applications require. In this article, 3D printing technology is used to its full potential to produce electrodes, which are modified into pH sensors and reference electrodes, with shapre‐driven additional functionality for monitoring ammonia recovery in a bioreactor. The 3D‐printed electrodes are modified with a layer of iridium oxide to be turned into pH sensitive devices. Their characterization showed their characteristic super‐Nernstian response (−77 ± 0.2 mV pH −1 ), high reproducibility (RSD < 5%) between sensors and repeatability (RSD < 2%) between measurements. Moreover, the sensors are stable for at least 20 days and tunable in length. All of this results in the sensors being built into a functional shape and tested to monitor the performance of an ammonia‐producing bioelectrochemical reactor.

AI summary

72% confidence

This paper develops 3D-printed iridium oxide (IrOx) pH sensors integrated into a modular platform for monitoring ammonia recovery in a bioelectrochemical reactor. The sensors achieve super-Nernstian response (−77 ± 0.2 mV/pH), high reproducibility (RSD < 5%), and stability for ≥20 days. The platform enables multi-point pH profiling in the recovery chamber where H⁺ and NH₄⁺ are converted to H₂ and NH₃ via gas-diffusion cathode, with recovered NH₃ acidified back to NH₄⁺ salts. The work demonstrates that 3D printing enables custom sensor geometries and integration with reactor-specific components, addressing a gap in electrochemical sensing tools for ammonia recovery systems.

Key insights

  • Performance IrOx electrodes of different lengths (64–118 mm) maintain consistent sensitivity (79–80.1 mV/pH) despite 1.4% variation, proving that 3D-printed device electrical resistance does not limit pH sensor performance across the design range.

    So what. Designers can use longer electrodes for spatial profiling in the recovery chamber without recalibrating or sacrificing sensitivity, enabling asymmetry mapping of cathode performance.

  • Threshold Sensor sensitivity decreases 9–12% over 20 days, but the rate of decay halves after day 5, stabilizing sufficiently for 15+ day operation and week-long experiments with minimal replacement.

    So what. Operators can plan sensor replacement intervals around 15 days for continuous monitoring without frequent maintenance, but should expect initial drift in the first week.

  • Performance 3D-printed Ag/AgCl reference electrode shows only 2.9 ± 0.3% difference in sensitivity compared to commercial reference despite 225 ± 2 mV potential shift, confirming interchangeability and reversibility.

    So what. Fully 3D-printed sensor platforms can replace commercial reference electrodes without recalibration, enabling true miniaturization and integration into custom reactor geometries.

  • Trade-off Electrode width must be increased to 1.2 mm to prevent contact with reactor cathode and mechanical failure under plate pressure, revealing a critical design constraint for integration.

    So what. When integrating 3D-printed sensors into plate reactors, account for mechanical tolerance and electrical isolation by increasing non-conductive shell width; test fit before full assembly.

  • Mechanism pH response time varies inversely with pH extremity: <15 s at pH < 5 and pH > 10, but 50–120 s in the neutral range (5 < pH < 10), indicating slower kinetics in buffered regions.

    So what. For real-time monitoring of ammonia recovery (where pH rises from ~1 toward neutral as acid is consumed), expect slower sensor response in the critical mid-range; use averaging or predictive algorithms.

Opportunity areas6

Specific gaps in this paper, with an experiment that would close each.

  • High No data on pH sensor performance in the presence of dissolved ammonia (NH₃) or ammonium (NH₄⁺) at the concentrations expected in the recovery chamber (likely > 0.1 M after acid neutralization). Interference from these species on IrOx response is not characterized.

    Section 2.4 states the recovery chamber is filled with 1.0 M H₂SO₄ and ammonia is generated and diffuses into it, but all calibrations (Figures 3–7) were performed in phosphate buffer or pure acid/base titrations without NH₃/NH₄⁺ present. The paper does not report whether high ammonium concentration affects sensor sensitivity or response time.

    Try. Calibrate sensors in 1.0 M H₂SO₄ spiked with 0.1–1.0 M NH₄⁺ or NH₃ and compare sensitivity and response time to calibrations in buffer alone; measure potential drift over 24 h in static high-ammonium solution.

  • High No measurement of actual pH evolution or ammonia recovery rate during bioelectrochemical reactor operation. The paper states sensors were 'placed in the recovery chamber' and 'tracked the operativity of the biological reactor by indirectly measuring the amount of ammonium recovered' but provides no time-series data, recovery efficiency, or correlation between pH rise and NH₄⁺ accumulation.

    Section 2.4 ends abruptly: 'With such a sensor distribution, a clearer picture of the asymmetry of the cathode can be studied... Moreover, they tracked the operativity of the biological reactor by indirectly measuring the amount of ammonium recovered. The recovery chamber is filled with 1.0 M H₂SO₄...' and then the text is cut off ('cou'). No results from reactor operation are presented.

    Try. Run a complete bioelectrochemical ammonia recovery cycle with the 3D-printed sensor platform; record pH at 5 different heights/positions over 24–72 h; measure final NH₄⁺ concentration by ion chromatography or titration; correlate pH rise rate to current density and cathode potential.

  • High No investigation of sensor performance under the electrochemical conditions of the bioelectrochemical reactor (applied cathode potential, current density, presence of biofilm). All characterization was done at open-circuit potential (OCP) in abiotic solutions.

    Section 2.1 states 'Measurements were performed under continuous stirring and in OCP conditions so a Nernstian response can be observed.' The reactor operates under applied potential (cathode for ammonia generation), which may cause electrochemical noise, fouling, or potential interference with the sensor's reference electrode.

    Try. Calibrate and measure pH response of 3D-printed sensors in a three-electrode cell with applied cathode potential (−0.5 to −1.0 V vs Ag/AgCl) and current density (10–100 mA/cm²) typical of ammonia recovery; assess potential drift and noise; test in presence of biofilm or bacterial culture.

  • Medium No data on sensor performance in the presence of sulfate (SO₄²⁻) at 1.0 M concentration (from H₂SO₄ in recovery chamber). Potential interference from high ionic strength or specific anion adsorption on IrOx is not addressed.

    Section 2.4 specifies 'The recovery chamber is filled with 1.0 M H₂SO₄' but all calibrations used phosphate buffer (1 mM) or dilute acid/base. High sulfate concentration may affect electrode surface chemistry, double-layer capacitance, or proton activity coefficient.

    Try. Calibrate sensors in 1.0 M H₂SO₄ (matching reactor conditions) and compare sensitivity and response time to calibrations in phosphate buffer; measure potential drift in static 1.0 M H₂SO₄ over 7 days.

  • Medium No characterization of sensor behavior at pH < 2 or pH > 12, despite the paper mentioning that conventional glass electrodes have poor accuracy at pH < 1 and strong Na⁺ interference at pH > 12. The operating range of the 3D-printed sensors in extreme conditions is not defined.

    Figure 3A shows calibration from pH ~2 to pH ~12, but the paper does not explicitly state the validated operating range or discuss performance at the extremes. The recovery chamber starts at pH ~0 (1.0 M H₂SO₄) and rises as ammonia is absorbed; the lower bound is critical but untested.

    Try. Calibrate sensors in 1.0 M H₂SO₄ (pH ~0) up to pH 2, and separately in pH 11–13 solutions; report sensitivity, response time, and hysteresis at each extreme; define the validated operating range.

  • Low No comparison of 3D-printed sensor cost, fabrication time, or material consumption versus commercial pH electrodes. The economic and practical advantages of the 3D-printing approach are not quantified.

    The paper emphasizes that 'conventional glass-type pH meters... suffer from drawbacks such as high fabrication cost, difficulty of miniaturization, glass-electrode brittleness' and that 'there is a significant lack of research into pH sensors dedicated to this field,' but does not provide cost or time data for the 3D-printed alternative.

    Try. Document material cost (PLA filament, IrOx precursor, Ag ink, PVB membrane), fabrication time (printing, activation, electrodeposition, assembly), and labor per sensor; compare to commercial electrode cost and lead time; calculate cost per sensor and cost per measurement over 20-day lifetime.

Generated by MESSAI extraction pipeline · review against source PDF

02

Extracted values

Every value pulled from the paper, with its source, so each number can be checked against the text.

n = 25 extracted values

Select a citation to see the quote, page and confidence behind a value.

1 of 25 extracted rows modelable

  • 19 not canonicalised (no canonical slug)
  • 5 no SI value
ParameterValueUnitConfidenceSourceNotesModelable
External Load11.2kΩ90%
extracted
"Printed electrodes of 64 mm of length (N = 6) had an average electrical resistan…" · substrate=phosphate buffer saline solutionModelable
Sensitivity decrease Sensor 2 after 20 days12%90%
extracted
"with Sensor 2 following closely with 12 3% after the same period" · substrate=Milli-Q water or pH 4.01 bufferExcluded: not canonicalised (no canonical slug)
Sensitivity decrease Sensor 1 after 20 days9%90%
extracted
"For Sensor 1, a decrease of 9 2% in sensitivity was accounted for after 20 days…" · substrate=Milli-Q water or pH 4.01 bufferExcluded: not canonicalised (no canonical slug)
Electrode 3 Slope72mV90%
extracted
"Electrode 3: Slope [mV] 72 3" · substrate=phosphate buffer saline solutionExcluded: not canonicalised (no canonical slug)
Electrode 1 Slope70mV90%
extracted
"Electrode 1: Slope [mV] 70 3" · substrate=phosphate buffer saline solutionExcluded: not canonicalised (no canonical slug)
Electrode 2 Slope72mV90%
extracted
"Electrode 2: Slope [mV] 72 4" · substrate=phosphate buffer saline solutionExcluded: not canonicalised (no canonical slug)
Sensitivity alkaline to acidic direction78.5mV/pH90%
extracted
"Going from an alkaline to an acidic pH resulted in 78.5 0.7 mV pH1" · substrate=phosphate buffer saline solutionExcluded: not canonicalised (no canonical slug)
RSD between slopes4.7%90%
extracted
"a 4.7% RSD between slopes and 1.8% RSD between intercepts are calculated by erro…" · substrate=phosphate buffer saline solutionExcluded: not canonicalised (no canonical slug)
Potential shift between commercial and 3D-reference electrodes225mV90%
extracted
"Between both RE, commercial and 3D-reference, the potentials shifted 225 2 mV" · substrate=phosphate buffer saline solutionExcluded: not canonicalised (no canonical slug)
Sensitivity with 3D-printed RE79.4mV/pH90%
extracted
"s= (79.4 0.3) mV/pH" · substrate=phosphate buffer saline solutionExcluded: not canonicalised (no canonical slug)
Response time at 5 < pH < 1050s90%
extracted
"It's in the intermediate range, 5 < pH < 10, where response was slower, between …" · substrate=phosphate buffer saline solutionExcluded: not canonicalised (no canonical slug)
Sensor Sensitivity80.1mV/pH90%
extracted
"for 118 mm it was 80.1 0.8 mV pH1" · substrate=phosphate buffer saline solutionExcluded: no SI value
Sensor Sensitivity77mV/pH90%
extracted
"Sensitivity average between three independent calibrations of a single sensor wa…" · substrate=phosphate buffer saline solutionExcluded: no SI value
Sensitivity with commercial RE81.8mV/pH90%
extracted
"s= (81.80.3) mV/pH" · substrate=phosphate buffer saline solutionExcluded: not canonicalised (no canonical slug)
Sensor Sensitivity79mV/pH90%
extracted
"For electrodes of 77.5 mm (Figure 6) a sensitivity of 79 1 mV pH1 was achieved" · substrate=phosphate buffer saline solutionExcluded: no SI value
Difference in sensitivities2.9%90%
extracted
"the difference in sensitivities between them was only 2.9 0.3%" · substrate=phosphate buffer saline solutionExcluded: not canonicalised (no canonical slug)
RSD between intercepts1.8%90%
extracted
"a 4.7% RSD between slopes and 1.8% RSD between intercepts are calculated by erro…" · substrate=phosphate buffer saline solutionExcluded: not canonicalised (no canonical slug)
Response time at pH < 5 and pH > 1015s90%
extracted
"At pH < 5 and pH > 10 response times were lower than 15 s." · substrate=phosphate buffer saline solutionExcluded: not canonicalised (no canonical slug)
Sensor Sensitivity80mV/pH90%
extracted
"for 104.5 mm of 80 1 mV pH1 and for 118 mm it was 80.1 0.8 mV pH1" · substrate=phosphate buffer saline solutionExcluded: no SI value
Sensitivity acidic to alkaline direction79mV/pH90%
extracted
"in the reverse order it was 79.0 1.0 mV pH1" · substrate=phosphate buffer saline solutionExcluded: not canonicalised (no canonical slug)
RSD Sensor 1 at pH 41.2%90%
extracted
"Sensor 1 measuring at pH 4 had the largest RSD, at 1.2%" · substrate=pH buffer solutionsExcluded: not canonicalised (no canonical slug)
Days until first ammonia detection at Sensor 21day90%
extracted
"For Sensor 2, only one day after initiating the operation of the reactor there w…" · substrate=1.0 M H2SO4 recovery solution with ammonia diffusionExcluded: not canonicalised (no canonical slug)
Days until first ammonia detection at Sensor 12day90%
extracted
"while Sensor 1 didn't notice until the second day" · substrate=1.0 M H2SO4 recovery solution with ammonia diffusionExcluded: not canonicalised (no canonical slug)
Sensor Sensitivity80mV/pH90%
extracted
"The normal response for this type of electrodes is around 80 2 mV/pH for the 64…" · substrate=phosphate buffer saline solutionExcluded: no SI value
Ammonia-related acid consumption0.4mol90%
extracted
"This revealed that ammonium recovering was fast, consuming about 0.4 Hþ mols in …" · substrate=1.0 M H2SO4 recovery solution with ammonia diffusionExcluded: not canonicalised (no canonical slug)

Methodology

What was the system and how was it run

Methodology

Methodology

3D-printed carbon electrodes (PLA filament, 25 mm² projected area, 36 mm² usable area) were activated in 1.0 M NaOH by chronoamperometry, then modified with iridium oxide via cyclic voltammetry electrodeposition. Sensors were characterized in phosphate buffer (1 mM, pH 7.4) by titration with HCl/NaOH under OCP conditions, with simultaneous measurement by commercial pH electrode. Reproducibility was assessed across three independently fabricated sensors (three-point buffer calibrations); repeatability was tested by repeated measurements at pH 4 and 9; long-term stability was monitored over 20 days with different storage conditions. A 3D-printed Ag/AgCl reference electrode was fabricated and compared to commercial reference. Electrodes of four different lengths (64–118 mm) were tested to assess sensitivity consistency. The integrated platform (five working electrodes + two reference electrodes in a 98 × 98 × 19.8 mm checkered spacer) was assembled into an ammonia-recovery bioelectrochemical reactor with a gas-diffusion cathode and recovery chamber filled with 1.0 M H₂SO₄.

System configuration

FieldValue
reactor typeMERC
chamber config2-chamber

Operating conditions

5 runs extracted
RunT°CpHanpHcatSubstrate[S]mg/LHRThModeR_extΩV_appVInoculum
v2-2df4e451d653cc5d———phosphate buffer saline solution———11200——
v2-a15b656a95cdac97———Milli-Q water or pH 4.01 buffer——————
v2-aab1ae703455d11e———pH buffer solutions——————
v2-c307aa59ea8b6c43———1.0 M H2SO4 recovery solution with ammonia diffusion——————
v2-c3a9b1b16882dd28———phosphate buffer saline solution——————

Pipeline

Raw extractions, artifacts, quality signal

Extraction quality

honesty signal

Calibrate your trust in the extracted content against these signals — they're how the pipeline rated its own output for this paper.

AI confidence

72%

Model self-reported confidence

Mean parameter confidence

90%

Average confidence across PaperParameter rows surfaced above

review against source PDF before citing · scores 0–1

03

Chart explorer

Move the inputs this paper used and see how the predicted outcome shifts.

04

Explore model

The reactor in 3D, with flow you can switch on and off.

05

AI insights

Contradictions, gaps and opportunities found by comparing this paper with the corpus.

06

Compare similar papers

The nearest papers by system, materials and organisms, with their outcomes side by side.

07

Source PDF

The paper as published.

No 3D model is mapped to this paper yet. The values extracted from it are listed under Extracted values.

25
Extracted values
2025
Year

Keywords

  • Fabrication
  • Reproducibility
  • Repeatability
  • 3D printing
  • Electrode
  • Layer (electronics)

Cite

Franc Paré, Mingyue Pan, Oscar Guerrero‐Sodric et al. (2025). 3D Printing as an Adaptive Tool for Sensor Fabrication Applied through pH Monitoring. ChemElectroChem. https://doi.org/10.1002/celc.202500264