Cyclic voltammetry, square wave voltammetry or electrochemical impedance spectroscopy? Interrogating electrochemical approaches for the determination of electron transfer rates of immobilized redox proteins
Ferd Pilz, Patrycja Kielb
Reported parameters
No values extracted from this paper yet.
No 3D model is mapped to this paper yet. Parameter ranges above still place reported values on the literature distribution.
Abstract
In this article, we cross-examine three well-established electrochemical approaches, namely cyclic voltammetry (CV), cyclic square-wave voltammetry (SWV) and electrochemical impedance spectroscopy (EIS) to dissect the electron transfer (ET) rate of electrostatically immobilized cytochrome c on Ag electrodes. A detailed analysis supported by simulations of redox transition provided three distinct values for the heterogeneous electron transfer (HET) rate constant of cyt c interfaced on COOH-terminated C 10 -long alkanethiol, i.e., k HET = 47.8 (±2,91) s -1 in CV, k HET = 64.8 (±1,27) s -1 in SWV, and k HET = 26.5 s -1 in EIS. We discuss the obtained discrepancies obtained from electrochemical methods and compare them with the data from spectro-electrochemical experiments. A comprehensive selection list is created from which the most applicable approach can be chosen for studying proteins of interest. CV is most applicable to study the interfaced proteins exhibiting k HET of ca. 0.5 - 70 s -1 , SWV is suitable for a broader range of k HET of 5 - 120 s -1 and EIS for k HET of 0.5 to 5 s -1 if alkanethiols are used as immobilization strategy.
Keywords
Identifiers
- PubMed
- 37424628
- Journal
- BBA Advances
- Year
- 2023