Extraction

Reported parameters

No values extracted from this paper yet.

No 3D model is mapped to this paper yet. Parameter ranges above still place reported values on the literature distribution.

Abstract

Glutathione is an important biological molecule which can be an indicator of numerous diseases. A method for self-powered detection of glutathione levels in solution has been developed using an enzymatic biofuel cell. The device consists of a glucose oxidase anode and a bilirubin oxidase cathode. For the detection of glutathione, the inhibition of bilirubin oxidase leads to a measurable decrease in current and power output. The reported method has a detection limit of 0.043 mM and a linear range up to 1.7 mM. Being able to detect a range of concentrations can be useful in evaluating a patient's health. This method has the potential to be implemented as a quick, low-cost alternative to previously reported methods.

Keywords

Bilirubin oxidaseGlutathioneDetection limitBiosensorLinear rangeGlucose oxidase

Identifiers

PubMed
32899114
Journal
Biosensors
Year
2020