Electrochemical Biosensor for Rapid and Sensitive Detection of Magnetically Extracted Bacterial Pathogens
Emma B. Setterington, Evangelyn C. Alocilja
AI summary
100% confidenceMES research in electrode, membrane. Biological defense and security applications demand rapid, sensitive detection of bacterial ...
Generated by MESSAI extraction pipeline · review against source PDF
No 3D model is mapped to this paper yet. Parameter ranges above still place reported values on the literature distribution.
Abstract
Biological defense and security applications demand rapid, sensitive detection of bacterial pathogens. This work presents a novel qualitative electrochemical detection technique which is applied to two representative bacterial pathogens, Bacillus cereus (as a surrogate for B. anthracis) and Escherichia coli O157:H7, resulting in detection limits of 40 CFU/mL and 6 CFU/mL, respectively, from pure culture. Cyclic voltammetry is combined with immunomagnetic separation in a rapid method requiring approximately 1 h for presumptive positive/negative results. An immunofunctionalized magnetic/polyaniline core/shell nano-particle (c/sNP) is employed to extract target cells from the sample solution and magnetically position them on a screen-printed carbon electrode (SPCE) sensor. The presence of target cells significantly inhibits current flow between the electrically active c/sNPs and SPCE. This method has the potential to be adapted for a wide variety of target organisms and sample matrices, and to become a fully portable system for routine monitoring or emergency detection of bacterial pathogens.
Keywords
Identifiers
- Journal
- Biosensors
- Year
- 2012