Researchers developed two MFC sensors, MFC-A and MFC-B, using sodium acetate and glucose as substrates, respectively, and found that MFC-A outperformed MFC-B in detecting sodium acetate and glucose.

Generated by MESSAI extraction pipeline · review against source PDF

Generic MFCRepresentative model
Click to animate flow
loading 3D model…

Representative MFC — matched on the paper’s system type only, not its reactor or geometry.

Extraction

Reported parameters

2 extracted values

View extracted values

Open in lab for full controls, parameter editing, and template overlays.

Open in lab →

What they did

System
MFC

What worked

No outcome metrics extracted yet.

Abstract

Most microbial fuel cell (MFC) sensors only focus on the detection of mixed solutions with respect to the chemical oxygen demand (COD) or toxicity; however, the concentrations of the individual analytes in a mixed solution have rarely been studied. Herein, we developed two types of MFC sensors, adapted with sodium acetate (MFC-A) and glucose (MFC-B) as organic substrates in the startup period. An evident difference in the sensor sensitivities (the slope value of the linear-regression curve) was observed between MFC-A and MFC-B. MFC-A exhibited a superior performance compared with MFC-B in the detection of sodium acetate (4868.9 vs. 2202 mV/(g/L), respectively) and glucose (3895.5 vs. 3192.9 mV/(g/L), respectively). To further compare these two MFC sensors, the electrochemical performances were evaluated, and MFC-A exhibited a higher output voltage and power density (593.76 mV and 129.81 ± 4.10 mW/m², respectively) than MFC-B (484.08 mV and 116.21 ± 1.81 mW/m², respectively). Confocal laser scanning microscopy (CLSM) and microbial-community analysis were also performed, and the results showed a richer anode biomass of MFC-A in comparison with MFC-B. By utilizing the different sensitivities of the two MFC sensors towards sodium acetate and glucose, we proposed and verified a novel method for a simultaneous test on the individual concentrations of sodium acetate and glucose in a mixed solution. Linear equations of the two variables (concentrations of sodium acetate and glucose) were formulated. The linear equations were solved according to the output voltages of the two MFC sensors, and the solutions showed a satisfactory accuracy with regard to sodium acetate and glucose (relative error less than 20%).

Key findings

  • MFC-A exhibited superior performance in detecting sodium acetate (4868.9 vs. 2202 mV/(g/L))
  • MFC-A also outperformed MFC-B in detecting glucose (3895.5 vs. 3192.9 mV/(g/L))
  • MFC-A had higher output voltage and power density (593.76 mV and 129.81 ± 4.10 mW/m²) compared to MFC-B (484.08 mV and 116.21 ± 1.81 mW/m²)

Keywords

Microbial fuel cellSodium acetateAnalyteSodiumEthyl acetateBiosensor