Publikation · MFC

Microfluidic Bioelectrochemical Cell Platform for the Study of Extracellular Electron Uptake in Microbes

01

Zusammenfassung

Was die Publikation berichtet, in ihren eigenen Worten.

Interaktives 3D-ModellKI-angereichert· 90 %

Zusammenfassung

Miniaturization of measurement systems offers several advantages, including reduced sample and reagent volumes, improved control over experimental conditions, and the ability to multiplex complementary measurement modalities, thereby enabling new types of studies in microbial electrochemistry. We present a scalable glass-based microfluidic bioelectrochemical cell (µ-BEC) platform for multiplexed investigations of microbial extracellular electron uptake (EEU). The platform integrates eight independently addressable three-electrode cells in a 2×4 array, with transparent working electrodes that support simultaneous electrochemical analysis and optical imaging. Using Rhodopseudomonas palustris TIE-1 as a model phototroph, we measured EEU activity under light-dark cycling. Microfluidic flow was used to selectively remove planktonic cells, enabling isolation of the electron uptake signal associated with surface attached cells. These results demonstrate the µ-BEC as a robust and adaptable platform for probing microbial electron transfer, with broad potential for high-throughput and multimodal studies.

KI-Zusammenfassung

90 % Konfidenz

International BES research in biotechnology with focus on biotechnology, electrode. Miniaturization of measurement systems offers several advantages, including reduced sample and reagent volumes, ...

Erstellt von der MESSAI-Extraktionspipeline · mit dem Quell-PDF abgleichen

02

Extrahierte Werte

Jeder aus der Publikation entnommene Wert mit seiner Quelle, damit sich jede Zahl am Text prüfen lässt.

n = 43 extrahierte Werte

Wählen Sie eine Quellenangabe, um Zitat, Seite und Konfidenz hinter einem Wert zu sehen.

14 von 43 extrahierten Zeilen modellierbar

  • 19 nicht kanonisiert (kein kanonischer Slug)
  • 3 kein SI-Wert
  • 7 nicht als modellierbar markiert
ParameterWertEinheitKonfidenzQuelleAnmerkungenModellierbar
Pseudoreference electrode exposed surface area1,29mm²95 %
device design specification
"The exposed surface areas of the working, counter, and pseudoreference electrode…" · electrode_material=silverAusgeschlossen: nicht kanonisiert (kein kanonischer Slug)
Electrode Surface Area1,96mm²95 %
device design specification
"The exposed surface areas of the working, counter, and pseudoreference electrode…" · electrode_material=indium tin oxide (ITO)Modellierbar
Electrode Surface Area5,93mm²95 %
device design specification
"The exposed surface areas of the working, counter, and pseudoreference electrode…" · electrode_material=goldModellierbar
Channel Width500µm95 %
device design specification
"To enable efficient removal of planktonic cells during microbial experiments, ea…"Modellierbar
Channel Depth100µm95 %
device design specification
"These electrodes are enclosed within 4-mm diameter, 100-µm deep wells isotropica…"Modellierbar
Microfluidic well diameter4mm95 %
device design specification
"These electrodes are enclosed within 4-mm diameter, 100-µm deep wells isotropica…"Ausgeschlossen: nicht kanonisiert (kein kanonischer Slug)
Standard Potential214mV90 %
cyclic voltammetry calibration
"Commercial Ag/AgCl/3 M KCl reference electrodes exhibited an average half-wave p…" · substrate=1 mM ferri/ferrocyanide in 1 M KClAusgeschlossen: nicht als modellierbar markiert
Pseudoreference electrode final potential131mV90 %
open circuit potential measurement
"and concluded at 131 mV (standard deviation: 8.82 mV, CV: 6.7%), corresponding t…" · substrate=freshwater mediaAusgeschlossen: nicht kanonisiert (kein kanonischer Slug)
Pseudoreference electrode net potential change-36mV90 %
open circuit potential measurement
"corresponding to a net potential change of –36 mV over the 120-hour testing peri…" · substrate=freshwater mediaAusgeschlossen: nicht kanonisiert (kein kanonischer Slug)
Pseudoreference electrode potential drift-0,3mV h⁻¹90 %
open circuit potential measurement
"The pseudoreference electrodes exhibited an average potential drift of –0.30 mV …" · substrate=freshwater mediaAusgeschlossen: nicht kanonisiert (kein kanonischer Slug)
Pseudoreference electrode initial potential167mV90 %
open circuit potential measurement
"The electrodes began at an average potential of 167 mV (standard deviation: 4.54…" · substrate=freshwater mediaAusgeschlossen: nicht kanonisiert (kein kanonischer Slug)
Pseudoreference electrode potential offset versus Ag/AgCl/3 M KCl-40mV90 %
cyclic voltammetry calibration
"corresponding to an average potential offset of –40 mV versus Ag/AgCl/3 M KCl" · substrate=1 mM ferri/ferrocyanide in 1 M KClAusgeschlossen: nicht kanonisiert (kein kanonischer Slug)
Standard Potential174mV90 %
cyclic voltammetry calibration
"microfabricated Ag/AgCl/NafionTM pseudoreference electrodes displayed an average…" · substrate=1 mM ferri/ferrocyanide in 1 M KClAusgeschlossen: nicht als modellierbar markiert
Anodic peak potential at 25 mV/s252mV85 %
cyclic voltammetry
"Scan Rate: 25 mV/s, Anodic Peak Potential [mV] Average: 252" · substrate=1 mM ferri/ferrocyanide in 1 M KClAusgeschlossen: nicht kanonisiert (kein kanonischer Slug)
Peak Current4,67µA85 %
cyclic voltammetry
"Scan Rate: 20 mV/s, Anodic Peak Current [µA] Average: 4.67" · substrate=1 mM ferri/ferrocyanide in 1 M KClModellierbar
Peak Current3,93µA85 %
cyclic voltammetry
"Scan Rate: 15 mV/s, Cathodic Peak Potential [µA] Average: 3.93" · substrate=1 mM ferri/ferrocyanide in 1 M KClModellierbar
Peak Current2,15µA85 %
cyclic voltammetry
"Scan Rate: 5 mV/s, Cathodic Peak Potential [µA] Average: 2.15" · substrate=1 mM ferri/ferrocyanide in 1 M KClModellierbar
Peak Current3,23µA85 %
cyclic voltammetry
"Scan Rate: 10 mV/s, Cathodic Peak Potential [µA] Average: 3.23" · substrate=1 mM ferri/ferrocyanide in 1 M KClModellierbar
Anodic peak potential at 20 mV/s252mV85 %
cyclic voltammetry
"Scan Rate: 20 mV/s, Anodic Peak Potential [mV] Average: 252" · substrate=1 mM ferri/ferrocyanide in 1 M KClAusgeschlossen: nicht kanonisiert (kein kanonischer Slug)
Anodic peak potential at 15 mV/s254mV85 %
cyclic voltammetry
"Scan Rate: 15 mV/s, Anodic Peak Potential [mV] Average: 254" · substrate=1 mM ferri/ferrocyanide in 1 M KClAusgeschlossen: nicht kanonisiert (kein kanonischer Slug)
Cathodic R-squared value1—85 %
randles-ševčík analysis
"Anodic and cathodic peak currents exhibited a linear relationship with the squar…" · substrate=1 mM ferri/ferrocyanide in 1 M KClAusgeschlossen: nicht kanonisiert (kein kanonischer Slug)
Peak Current4,12µA85 %
cyclic voltammetry
"Scan Rate: 15 mV/s, Anodic Peak Current [µA] Average: 4.12" · substrate=1 mM ferri/ferrocyanide in 1 M KClModellierbar
Anodic R-squared value0,999—85 %
randles-ševčík analysis
"Anodic and cathodic peak currents exhibited a linear relationship with the squar…" · substrate=1 mM ferri/ferrocyanide in 1 M KClAusgeschlossen: nicht kanonisiert (kein kanonischer Slug)
Anodic peak potential at 10 mV/s254mV85 %
cyclic voltammetry
"Scan Rate: 10 mV/s, Anodic Peak Potential [mV] Average: 254" · substrate=1 mM ferri/ferrocyanide in 1 M KClAusgeschlossen: nicht kanonisiert (kein kanonischer Slug)
Anodic peak potential at 5 mV/s252mV85 %
cyclic voltammetry
"Scan Rate: 5 mV/s, Anodic Peak Potential [mV] Average: 252" · substrate=1 mM ferri/ferrocyanide in 1 M KClAusgeschlossen: nicht kanonisiert (kein kanonischer Slug)
25 von 43 angezeigt

Methodik

Welches System und wie es betrieben wurde

Methodik

Methodology

A scalable glass-based microfluidic bioelectrochemical cell (µ-BEC) platform was developed to study microbial extracellular electron uptake (EEU) by integrating eight independently addressable three-electrode cells in a 2×4 array.

Biologische Parameter

FeldWert
dominant species["Rhodopseudomonas"]

Betriebsbedingungen

6 Läufe extrahiert
LaufT°CpHAnolytpHKatholytSubstrat[S]mg/LHRThModusR_extΩV_appVInokulum
v2-03e6b022e08bfe4630——ferri_ferrocyanide_in_kcl——————
v2-3f895831069b849130——freshwater medium——steady_state—0.1—
v2-6215ab02fda8a19d30——freshwater medium——startup—0.1—
v2-7abfc6466b5a1799———freshwater media——————
v2-ce7dcd6e9520abc8———1 mM ferri/ferrocyanide in 1 M KCl——————
v2-f586e9af22985397——————————

Extraktionspipeline

Rohextraktionen, Artefakte, Qualitätssignal

Extraktionsqualität

Ehrlichkeitssignal

Bemessen Sie Ihr Vertrauen in die extrahierten Inhalte an diesen Signalen – so hat die Pipeline ihre eigene Ausgabe für diese Publikation bewertet.

Parametervollständigkeit

13 %

Anteil der erwarteten Parameter, die extrahiert wurden

Datenvollständigkeit

30 %

Gesamtabdeckung der Extraktion über alle Felder

KI-Konfidenz

90 %

Vom Modell selbst angegebene Konfidenz

Mittlere Parameterkonfidenz

86 %

Durchschnittliche Konfidenz der oben angezeigten PaperParameter-Zeilen

vor dem Zitieren mit dem Quell-PDF abgleichen · Werte 0–1

03

Diagramm-Explorer

Verändern Sie die Eingaben dieser Publikation und sehen Sie, wie sich das vorhergesagte Ergebnis verschiebt.

04

Modell erkunden

Der Reaktor in 3D, mit Strömung, die sich ein- und ausschalten lässt.

05

KI-Erkenntnisse

Widersprüche, Lücken und Chancen aus dem Vergleich dieser Publikation mit dem Korpus.

06

Ähnliche Publikationen vergleichen

Die nächsten Publikationen nach System, Materialien und Organismen, mit ihren Ergebnissen nebeneinander.

07

Quell-PDF

Die Publikation, wie sie erschienen ist.

3D-Modell wird geladen …

Repräsentatives Modell, generisches MFC

Repräsentatives Modell: MFC – zugeordnet nur anhand des Systemtyps der Publikation, nicht ihres Reaktors oder ihrer Geometrie.

MFC
System
43
Extrahierte Werte
2025
Jahr

In diesem Datensatz

Organismen, Parameter und Materialien, die diese Publikation berichtet. Jeder Eintrag verweist auf seine Seite.

Schlagwörter

  • Microfluidics
  • Extracellular
  • Cell
  • Cell biology
  • Nanotechnology
  • Biophysics

Zitieren

Andreea Stoica, Karthikeyan Rengasamy, Tahina Onina Ranaivoarisoa et al. (2025). Microfluidic Bioelectrochemical Cell Platform for the Study of Extracellular Electron Uptake in Microbes. Preprints.org. https://doi.org/10.20944/preprints202505.1879.v1

Zeitschrift
Preprints.org

Öffnen Sie das Modell im Labor, um alle Steuerelemente, die Parameterbearbeitung und Vorlagen-Overlays zu nutzen.

Im Labor öffnen